3.2 Principle of operation
LEVEL SWITCH 8111 is a point level sensor with tuning fork for level
detection.
It is designed for industrial use in all areas of process technology and
can be used in liquids.
Typical applications are overll and dry run protection. With a tuning
fork of only 40 mm length, LEVEL SWITCH 8111 can be also mounted
e.g. in pipelines from DN 32. The small tuning fork allows use in
vessels, tanks and pipes. Thanks to its simple and robust measuring
system, LEVEL SWITCH 8111 is virtually unaected by the chemical
and physical properties of the liquid.
It functions even under dicult conditions such as turbulence, air
bubbles, foam generation, buildup, strong external vibration or
changing products.
Fault monitoring
The electronics module of LEVEL SWITCH 8111 continuously
monitors via frequency evaluation the following criteria:
l Strong corrosion or damage on the tuning fork
l Loss of vibration
l Line break to the pie zo drive
If one of these faults is detected or if the power supply fails, the
electronics takes on a dened output st atus (safe condition).
The tuning fork is piezoelectrically energised and vibrates at its
mechanical resonance frequency of approx. 1200 Hz. The piezos are
xed mechanically and are hence not subject to temperature shock
limitations. The frequency changes when the tuning fork is covered by
the medium. This change is detected by the integrated electronics
module and converted into a switching command.
LEVEL SWITCH 8111 is a compact instrument, i.e. it can be operated
without external evaluation system. The integrated electronics
evaluates the level signal and outputs a switching signal. With this
switching signal, a connected device can be operated directly (e.g. a
warning system, a pump etc.).
The data for power supply are specied in chapter "Technical data".
3.3 Operation
The switching condition of LEVEL SWITCH 8111 can be checked
when the housing is closed (signal lamp). With the basic setting,
products with a density > 0.7 g/cm³ (0.025 lbs/in³) can be detected.
The instrument can be adapted if products with lower density should
be measured.
On the electronics module you will nd the following indicating and
adjustment elements:
Appli
cation area
Func
tional principle
Volta
ge supply
8 LEVEL SWITCH 8111 with NAMUR output
3 Product description
32048-EN-120418
Seitenansicht 7
1 2 3 4 5 6 7 8 9 10 11 12 13 ... 111 112

Kommentare zu diesen Handbüchern

Keine Kommentare